These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Heimlich valve for chest drainage.
    Author: Heimlich HJ.
    Journal: Med Instrum; 1983; 17(1):29-31. PubMed ID: 6843411.
    Abstract:
    The Heimlich chest drainage valve was developed so that the process of draining the pleural cavity could be accomplished in a safe, relatively simple, and efficient manner. Replacing the cumbersome underwater drainage bottle system, the Heimlich valve connects to chest tubing and allows fluid and air to pass in one direction only. The valve, which functions in any position, need never be clamped, and regulated suction can be attached to it if necessary. The valve drains into a plastic bag that can be held at any level, allowing the patient undergoing chest drainage to be ambulatory simply by carrying the bag. The construction and function of the valve is easily understood by medical and nursing staffs. It is presterilized, stored in a sterile package, and readily utilized on emergency vehicles and in the operating room.
    [Abstract] [Full Text] [Related] [New Search]