These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Multi-enzyme membrane electrodes for determination of creatinine and creatine in serum. Author: Tsuchida T, Yoda K. Journal: Clin Chem; 1983 Jan; 29(1):51-5. PubMed ID: 6848282. Abstract: An enzyme electrode system for the determination of creatinine and creatine was developed by utilizing three enzymes: creatinine amidohydrolase (CA), creatine amidinohydrolase (Cl), and sarcosine oxidase (SO). These enzymes were co-immobilized onto the porous side of a cellulose acetate membrane with asymmetric structure, which has selective permeability to hydrogen peroxide. Two kinds of multi-enzyme electrodes were constructed by combining a polarographic electrode for sensing hydrogen peroxide and an immobilized CA/Cl/SO membrane or Cl/SO membrane for creatinine plus creatine or creatine, respectively. The multi-enzyme electrodes responded linearly up to 100 mg of creatinine and creatine per liter in human serum. Response time was 20 s in the rate method and the detection limit was 1 mg/L. Only 25 microL of serum sample is required. Analytical recoveries, precisions, and correlations with the Jaffé method were excellent, and the multi-enzyme electrodes were sufficiently stable to perform more than 500 assays. No loss of activity of immobilized enzymes was observed after nine months of storage at 4 degrees C in air.[Abstract] [Full Text] [Related] [New Search]