These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Estimation of aluminium in biological fluids by atomic absorption spectrophotometry in a graphite oven (author's transl)].
    Author: Clavel JP, Jaudon MC, Galli A.
    Journal: Ann Biol Clin (Paris); 1978; 36(1):33-8. PubMed ID: 686518.
    Abstract:
    The possibilities of use of a graphite oven type HGA 74 (Perkin Elmer) for the estimation of aluminium in biological fluids by atomic absorption spectrophotometry are described, with a special study of the stages of drying, mineralisation, atomisation, flow of inert gas, requirement for correction of non-specific absorption, etc. Under fixed experimental conditions, the limit of sensitivity (absorption 1%) was 2.28.10.(-11) grams and the linearity was good up to 100 microgram/l. This method of analysis by direct treatment of the samples required only small volumes of the latter and reduces the risk of contamination.
    [Abstract] [Full Text] [Related] [New Search]