These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: In vitro analysis of self-shearing retentive pins.
    Author: Collard EW, Caputo AA, Standlee JP, Duncanson MG.
    Journal: J Prosthet Dent; 1981 Feb; 45(2):156-9. PubMed ID: 6937657.
    Abstract:
    Combining stress, analysis, microscopic examination, mechanical testing of the shear mechanism, and retention of the Reten Pin leads to the following conclusions: 1. The suggested 0.006 inch pin-channel mismatch induces high lateral and apical stresses. Microscopically, this was seen to correlate with injury to the dentin. 2. The degree of retention was increased by using a smaller pin-channel mismatch. This correlates with smaller stresses and reduced dentinal damage. 3. The shear mechanism acts in a uniform manner, with a relatively small variation from the mean. It is suggested that for the best results the manufacturer should supply larger twist drills and pins with a somewhat deeper self-shearing groove to minimize apical involvement during shearing of the handle from the pin.
    [Abstract] [Full Text] [Related] [New Search]