These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Flexible valinomycin electrodes for on-line determination of intravascular and myocardial K+. Author: Hill JL, Gettes LS, Lynch MR, Hebert NC. Journal: Am J Physiol; 1978 Oct; 235(4):H455-9. PubMed ID: 696888. Abstract: Flexible electrode systems capable of monitoring in vivo changes in venous and myocardial extracellular potassium activity were constructed using valinomycin-polyvinyl chloride matrix membrane and polyvinyl chloride tubing. Electrode impedance was 1--30 Momega, time constant 10--200 ms, drift less than 1 mV/h, and shelf life approximately 3 days (intramyocardial electrode) and approximately 6 wk (venous electrode). In vitro and in vivo accuracy were determined in 5 dogs and 18 pigs, anesthetized with sodium pentobarbital (35 mg/kg iv), with normal and elevated K+ levels. Serum and venous K+ concentration, determined using a K+ electrode with tip diameter of 1.5 mm, correlated well to serum K+ values determined by flame photometry (r = 0.997). Steady-state myocardial extracellular K+ concentration determined using double-barrel electrodes with total tip diameter of 0.6 mm also correlated well to the serum K+ concentration (r = 0.992). These electrode assemblies permit on-line, virtually instantaneous in vivo determination of intravascular and local myocardial extracellular K+ activity, a capability not previously available.[Abstract] [Full Text] [Related] [New Search]