These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Hemadsorption immunosorbent technique for determination of rubella immunoglobulin M antibody. Author: van der Logt JT, van Loon AM, van der Veen J. Journal: J Clin Microbiol; 1981 Mar; 13(3):410-5. PubMed ID: 7016893. Abstract: A highly specific and sensitive hemadsorption immunosorbent technique for measuring rubella immunoglobulin M (IgM) antibody is described. IgM from human sera was absorbed into anti-human IgM-coated wells in plates and rubella-specific IgM was detected by adding rubella virus hemagglutinin and a small quantity of sheep erythrocytes. Centrifugation of the plates facilitated reading of the test. Specific IgM-positive sera showed hemadsorption, whereas negative sera showed hemagglutination. Rheumatoid factor and rubella-specific IgG antibody did not interfere with the results. The test was clearly more sensitive than the solid-phase immunosorbent technique for detection of rubella IgM antibody by hemagglutination inhibition and at least as sensitive as the hemagglutination inhibition test on IgM fractions from a sucrose density gradient and the indirect immunofluorescence test for IgM antibody with absorbed serum. All of 40 sera from 17 rubella patients taken 4 to 49 days after the onset of rash were positive in the new test, with antibody titers ranging from 2,560 to 81,920 between 4 and 28 days. The test is reliable, practical, and suitable for general diagnostic use.[Abstract] [Full Text] [Related] [New Search]