These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: A technique of scatter and glare correction for videodensitometric studies in digital subtraction videoangiography.
    Author: Shaw CG, Ergun DL, Myerowitz PD, Van Lysel MS, Mistretta CA, Zarnstorff WC, Crummy AB.
    Journal: Radiology; 1982 Jan; 142(1):209-13. PubMed ID: 7031761.
    Abstract:
    The logarithmic amplification of video signals and the availability of data in digital form make digital subtraction videoangiography a suitable tool for videodensitometric estimation of physiological quantities. A system for this purpose was implemented with a digital video image processor. However, it was found that the radiation scattering and veiling glare present in the image-intensified video must be removed to make meaningful quantitations. An algorithm to make such a correction was developed and is presented. With this correction, the videodensitometry system was calibrated with phantoms and used to measure the left ventricular ejection fraction of a canine heart.
    [Abstract] [Full Text] [Related] [New Search]