These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: The influence of pressure swings on the output of plenum vaporizers. Author: Burm AG, Jorritsma FF, van Bijnen AW. Journal: Anaesthesist; 1982 May; 31(5):257-61. PubMed ID: 7048993. Abstract: The influence of pressure swings on the output of Dräger Vapor, Dräger Vapor 19, Cyprane Fluotec Mark 2, Cyprane Fluotec Mark 3 and Ohio halothane vaporizers was investigated. It is shown that significant deviations from the dial setting occur with all vaporizers if the swings are large enough. The sensitivity for pressure swings is largest at low dial settings and low carrier flow rates. The results indicate that swings, occurring during assisted or controlled intermittent positive pressure ventilation, can induce significant deviations at low carrier flow rates. The frequency and in-/expiration time ratio are relatively of minor importance.[Abstract] [Full Text] [Related] [New Search]