These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of renal blood flow by thermodilution method. Author: Leivestad T, Brodwall EK, Simonsen S. Journal: Scand J Clin Lab Invest; 1978 Sep; 38(5):495-9. PubMed ID: 705231. Abstract: The single bolus thermodilution method for measurement of renal vein blood flow was tested. In model experiments the thermodilution method was compared with graduated cylinder measurements over a flow range from 50 to 1050 ml/min. There was a good correlation between the two methods (r = 0.98) with a mean of differences of 5.2%. In eighteen patients measurements were performed in duplicate in thirty-one renal veins. Comparison was made between the first (x) and second (u) measurement--performed within 3 min. The correlation between the two was very good (r = 0.99; y = 1.03x - 11.48). In twelve patients bilateral renal vein blood flow measurements were performed simultaneous to blood flow measurement by PAH clearance. The correlation between total flow measured by thermodilution (y) and by the clearance method (x) was good (r = 0.98; y = 0.79x + 221). It is concluded that the thermodilution method requires catheterization of the renal veins, but is otherwise simple to perform, is inexpensive and gives reliable results. It is particularly advantageous when repeated measurements in the study of acute changes in renal haemodynamics is desirable.[Abstract] [Full Text] [Related] [New Search]