These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of urinary oxalate by ion chromatography: preliminary observation. Author: Mahle CJ, Menon M. Journal: J Urol; 1982 Jan; 127(1):159-62. PubMed ID: 7057492. Abstract: The technique of ion chromatography was applied to determine oxalate concentrations in unprocessed urine. The minimal detectable limits of oxalate was 1 microgram./ml. (11 mumole/liter), the within-run imprecision (Sw) was 2.3 per cent and the total imprecision (St) was 4.9 per cent. Ion chromatography can be used for the simple, accurate and direct measurement of urinary oxalate.[Abstract] [Full Text] [Related] [New Search]