These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Mechanical plethysmography and Doppler ultrasound. Diagnosis of deep-venous thrombosis. Author: Schroeder PJ, Dunn E. Journal: Arch Surg; 1982 Mar; 117(3):300-3. PubMed ID: 7065871. Abstract: Two hundred eighty-one patients were available for follow-up after undergoing bilateral lower-extremity venous laboratory evaluations for possible deep-venous thrombosis. Because of a recognized difficulty with such a clinical diagnosis, the use of Doppler ultrasound and mechanical plethysmography with air-filled cuffs was carefully studied. Eighty-three of the 281 patients had confirmatory results from venography or fibrinogen scan of the symptomatic or more-symptomatic leg, and these 83 patients were used to determine accuracy. The complete evaluation as performed was accurate in 70% of the patients with suspected deep-venous thrombosis. The Doppler examination alone was accurate in 71% of the patients, and the maximum venous outflow measurement was accurate in 67% of the patients. This study indicates a continuing need for venography or fibrinogen scanning for the diagnosis of deep-venous thrombosis.[Abstract] [Full Text] [Related] [New Search]