These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Interferent ion effects on the Orion SS-20 Calcium Ion Analyser. Author: Vadgama P, Mitchison J, Covington AK, Alberti KG. Journal: Clin Chim Acta; 1982 Mar 12; 119(3):249-56. PubMed ID: 7067123. Abstract: Selectivity studies were carried out on bis-dialkyl phosphate and neutral carrier sensors available with the Orion SS-20 Calcium Analyser. Neutral carrier sensors showed better selectivity; changes of H+, Na+, K+ and Mg2+ within the physiological range produced no significant error in Ca2+ measurement. The Nicolsky equation, used as the basis of these studies, did not fully account for electrode response to mixed electrolytes. The importance of a check on the selectivity of sensors is emphasised.[Abstract] [Full Text] [Related] [New Search]