  • Title: Direct 125I-radioligand assays for serum progesterone compared with assays involving extraction of serum.
    Author: Ratcliffe WA, Corrie JE, Dalziel AH, Macpherson JS.
    Journal: Clin Chem; 1982 Jun; 28(6):1314-8. PubMed ID: 7074938.
    Abstract:
    We compared two direct radioimmunoassays for progesterone in 50 microL of unextracted serum or plasma with assays involving extraction of serum. The direct assays include the use of either danazol at pH 7.4 or 8-anilino-1-naphthalenesulfonic acid at pH 4.0 to displace progesterone from serum binding-proteins. Progesterone is then assayed by using an antiserum to a progesterone 11 alpha hemisuccinyl conjugate and the radioligand 125I-labeled progesterone 11 alpha-glucuronyl tyramine, with separation by double-antibody techniques. Direct assays with either displacing agent gave good analytical recovery of progesterone added to human serum, and progesterone values for patients' specimens correlated well (r greater than 0.96) with results of assays involving extraction of serum. Precision was similar with each displacing agent over the working range 2.5-100 nmol/L and superior to that of extraction assays. We conclude that these direct assays of progesterone are analytically valid and more robust, precise, and technically convenient than many conventional methods involving extraction of serum.