These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Continuous-flow analysis for serum albumin, by use of the immediate reaction with bromcresol green.
    Author: Plesher CJ.
    Journal: Clin Chem; 1978 Nov; 24(11):2036-9. PubMed ID: 709843.
    Abstract:
    Albumin reacts with bromcresol green reagent almost immediately, other serum proteins more slowly. This phenomenon can be exploited to improve the specificity of this routine method for determining serum albumin. I describe a 30-s reaction as adapted to continuous-flow analyzers, with no interference from other serum proteins. The method shows linearity to 72 g/liter. Results of this method correlate very well with those by the Laurell "rocket" and manual (30-s) bromcresol green methods. This procedure gives a considerable improvement in specificity for continuous-flow analyzers as compared with current bromcresol green methods.
    [Abstract] [Full Text] [Related] [New Search]