These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Controlled-delivery vapor generator for animal exposures.
    Author: Decker JR, Moss OR, Kay BL.
    Journal: Am Ind Hyg Assoc J; 1982 Jun; 43(6):400-2. PubMed ID: 7113919.
    Abstract:
    A vapor generation system is described which provides long-term stable vapor concentrations for inhalation bioassay tests with laboratory animals. Liquid is pumped from a reservoir by a micrometering pump to the surface of a cylindrical vaporizer covered with a glass fiber wick. An 80-watt heater and temperature sensor, embedded in the cylinder, control the vaporizer temperature. Pump rates of 0.01 to 20 mL/min and surface temperatures greater than 100 degrees C can be achieved. To minimize condensation on delivery-tube walls, the vaporizer is positioned in the dilution air inlet tube leading directly into the 2-m3 exposure chamber. During exposure of rats, mice and rabbits to eleven different chemicals for periods ranging from 4 hours to over 400 daily 6-hour exposure periods, concentrations from 25 to 10,000 ppm were achieved with a dilution air flow of 280 L/min. Chamber concentrations were routinely maintained within 10% of target levels. Stable pump rates provide a method of calculating vapor concentration independent of chemical analysis. Long-term generation stability and ease of operation reduce labor requirements compared to previous generation techniques.
    [Abstract] [Full Text] [Related] [New Search]