These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Total automated method for determination of cholesterol concentration in high density lipoprotein by heparin-magnesium-albumin precipitation.
    Author: Tallet F, Raichvarg D, Lefèvre G, Canal J, Yonger J.
    Journal: Clin Chim Acta; 1982 Aug 18; 123(3):339-45. PubMed ID: 7116650.
    Abstract:
    We describe an automated procedure for determination of high-density-lipoprotein cholesterol in serum by use of heparin-MgCl2-albumin reagent. 'AutoAnalyzer' II equipment was used in which high-density lipoprotein was separated from precipitate containing very-low and low-density lipoproteins by filtering on-line across a cellulose acetate membrane. Cholesterol concentration was measured by the enzymic method of Allain. Good correlation was obtained using the automated method compared to the heparin-MgCl2-albumin (r=0.99) or sodium phosphotungstate-MgCl2 procedure (r=0.97). Contamination was minimal with a contamination coefficient lower than 2.5%. The method presented was linear (0 to 2.50 mmol/1) and could be applied to lipemic samples with a coefficient of variation better than 3%. This study showed thus that this new procedure was suitable for routine work with a good rate (60 samples/h) in almost any laboratory.
    [Abstract] [Full Text] [Related] [New Search]