Title: A method for the routine determination of aluminium in serum and water by flameless atomic absorption spectrometry. Author: Parkinson IS, Ward MK, Kerr DN. Journal: Clin Chim Acta; 1982 Oct 27; 125(2):125-33. PubMed ID: 7139954. Abstract: A simple but reliable method for the routine determination of aluminium in serum and water by flameless atomic absorption spectrometry is described. No preparatory procedures are required for water samples, although serum is mixed with a wetting agent (Triton X-100) to allow complete combustion of the samples and to improve analytical precision. Precautions to prevent contamination during sample handling are discussed and instrumental parameters are defined. The method has a sensitivity of 35.5 pg and detection limits of 2.3 micrograms Al/l for serum and 1.3 micrograms Al/l for water. The method was used to determine the aluminium concentration in serum of 46 normal subjects. The mean aluminium content was 7.3 micrograms/l (range 2--15 micrograms/l.[Abstract] [Full Text] [Related] [New Search]