These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of volatile halocarbons in raw and drinking water, human serum, and urine by electron capture GC.
    Author: Reunanen M, Kroneld R.
    Journal: J Chromatogr Sci; 1982 Oct; 20(10):449-54. PubMed ID: 7142355.
    Abstract:
    A method for quantitative determination of volatile halocarbons in raw and drinking water, human serum, and urine is presented. Samples of water, serum, or urine were extracted in a single step with petroleum ether, and the extracts are, without further purification, analysed by high resolution gas chromatography (GC) with electron capture detection (ECD). Both external and internal calibrations were used to standardize the analytical system. External calibration is done with pure halocarbons in purified well water. Internal calibration is accomplished by spiking serum and urine samples with reference halocarbons between each of two successive extractions and GC runs. 1-iodobutane is used as internal standard, both in calibrations and actual determinations.
    [Abstract] [Full Text] [Related] [New Search]