These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [Properties of small nebulizers and their suitability for aerosol therapy]. Author: Wolf M, Angerstein W. Journal: Z Erkr Atmungsorgane; 1982; 159(2):210-4. PubMed ID: 7168215. Abstract: The current methods of measurement for particle size spectra do not seem to be applicable for small Hand-Held Nebulizers without impairing the relations of the amounts of fine and coarse particles. Therefore a simple method was developed that before-hand allowed a rough determination of these amounts. Measurements on four Hand-Held Nebulizers, three of them of the rubber ball nozzle-type and one ultrasonic, revealed that 90-95% of the medicament was unable to pass a 15 cm long rippled aerosol hose of 25 mm diameter. Therefore most of the drug is transformed into coarse particles and the application of Hand-Held Nebulizers is justified only for the therapy of the upper airways and for diseases where a resorption of the medicament in the main bronchi will suffice.[Abstract] [Full Text] [Related] [New Search]