These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Non-invasive measurement of cardiac output during anaesthesia. An evaluation of the soluble gas uptake method. Author: Heneghan CP, Branthwaite MA. Journal: Br J Anaesth; 1981 Apr; 53(4):351-5. PubMed ID: 7225268. Abstract: A technique for measuring cardiac output which depends on the uptake of an inert soluble gas from the lungs has been evaluated during anesthesia. A respiratory mass spectrometer has been used to follow the concentrations of argon and freon-22 during passive rebreathing in anaesthetized patients before cardiopulmonary bypass. Values for cardiac output obtained with this technique are reproducible, but lower than those recorded using the direct Fick technique before and after the rebreathing manoeuvre. A reduction in cardiac output caused by vigorous rebreathing is the most likely explanation for the discrepancy and, although serial measurements of oxygen consumption may permit application of a correction factor, a method of measurement which causes significant haemodynamic disturbance cannot be recommended for widespread use.[Abstract] [Full Text] [Related] [New Search]