  • Title: Ionized calcium analyzer with a built-in pH correction.
    Author: Fogh-Andersen N.
    Journal: Clin Chem; 1981 Jul; 27(7):1264-7. PubMed ID: 7237795.
    Abstract:
    We describe a new semi-automated apparatus for simultaneously measuring the concentration of free calcium ion and of hydrogen ion (pH) at 37 degrees C. The sample volume is 110 microL. In addition to the actual values for these concentrations in the sample, the apparatus calculates the concentration of free calcium ion at pH 7.40. Mean values for serum from 51 fasting bedridden patients without calcium metabolic disorders and 64 fasting hospital employees were 1.192 and 1.232 mmol/L, respectively, with SD of 0.042 and 0.040 mmol/L, respectively. The within-series analytical SD was 12 mumol/L and the day-to-day SD of the pH-corrected concentration of free calcium ion was 21 mumol/L, as calculated from measurements made on a serum pool after equilibration with a CO2--air mixture. The mean dependency on pH as determined in 120 consecutive patients' sera equalled the built-in pH correction. The accuracy was evaluated by comparison with other calcium ion-selective electrodes.
    [Abstract] [Full Text] [Related] [New Search]