These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Simultaneous determination of imipramine, desipramine, and their 2-hydroxy metabolites in plasma by ion-pair reversed-phase high-performance liquid chromatography with amperometric detection.
    Author: Suckow RF, Cooper TB.
    Journal: J Pharm Sci; 1981 Mar; 70(3):257-61. PubMed ID: 7264887.
    Abstract:
    An ion-pair reversed-phase high-performance liquid chromatographic (HPLC) method, using an electrochemical detector, is presented for the simultaneous and rapid quantitation of imipramine desipramine, and their 2-hydroxylated metabolites in plasma. The drugs are extracted from 1 ml of plasma at pH 9.7 with ether, back-extracted into 0. M HCl, and reextracted into ether following alkalinization. An efficient electrochemical oxidation reaction at the detector electrode affords a low detection level of approximately 5 ng/ml in a mobile phase of acetonitrile--acetate buffer (40:60) containing 0.005 M heptanesulfonate. Patient data are presented as correlations between the plasma level of each hydroxy metabolite and its respective parent compound. The method is applicable to the laboratory experienced in HPLC.
    [Abstract] [Full Text] [Related] [New Search]