These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Spontaneous rupture of an intra-aortic balloon pump.
    Author: Vicente JR, Moreno E, Peñas L, Diaz A, Fernandez L, Loma-Osorio A.
    Journal: Intensive Care Med; 1981; 7(6):311-2. PubMed ID: 7328223.
    Abstract:
    Circulatory assistance with an intraaortic balloon pump is not free from risks and complications. Among the latter, rupture is extremely rare. We report a case of spontaneous rupture of an intraaortic balloon, which occurred after 29 h of counterpulsation in a patient with cardiogenic shock following an acute myocardial infarction. The probable cause of the rupture was the constant abrasion and eventual perforation of the balloon membrane by a sharp calcified aortic plaque.