These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Corrosion fatigue of surgical stainless steel in synthetic physiological solution.
    Author: Cahoon JR, Holte RN.
    Journal: J Biomed Mater Res; 1981 Mar; 15(2):137-45. PubMed ID: 7348709.
    Abstract:
    Fatigue tests conducted both in air and synthetic physiological solution show that the fatigue strength of surgical stainless steel in synthetic physiological solution is about 10% lower than the strength in air for a given endurance level. It is proposed that surgical stainless steel which is normally passive in physiological solution suffers corrosion fatigue because of susceptibility to crevice corrosion which occurs at extrusions and intrusions (crevices) on the surface thereby shortening the crack initiation time and the fatigue life.
    [Abstract] [Full Text] [Related] [New Search]