These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Automated differential leucocyte counters: an evaluation of the Hemalog D and A comparison with the Hematrak. II. Evaluation of performance on routine blood samples from hospital patients. Author: Bain BJ, Scott D, Scott TJ. Journal: Pathology; 1980 Jan; 12(1):101-9. PubMed ID: 7375140. Abstract: The performance of the Hemalog D on routine blood samples from patients was assessed over a 6 wk period, 500 samples being studied in detail. The performance of the Hematrak was assessed on 85 routine blood samples. It was found that in our hospital population at least 58% of blood samples counted by the Hemalog D would require a blood film to be examined. Occasional patients with haematological diseases had abnormalities on the blood film but had normal Hemalog 8 and Hemalog D printouts. The Hemalog D was unable to perform an accurate differential count on the blood from a large proportion of patients with uraemia, including dialysis patients. The Hematrak was imprecise because of the small number of cells counted; it was also found to be inaccurate if blood films were not made rapidly after the blood was taken.[Abstract] [Full Text] [Related] [New Search]