These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Erroneously high Dextrostix values caused by isopropyl alcohol. Author: Grazaitis DM, Sexson WR. Journal: Pediatrics; 1980 Aug; 66(2):221-3. PubMed ID: 7402806. Abstract: Glucose oxidase peroxidase chromogen reagent (Dextrostix) in combination with the Eyetone colorimeter has become increasingly popular in the rapid detection of hypoglycemic states in the newborn. Although the reliability of this system is well documented, there are several factors which can compromise the accuracy of the procedure. One such problem is the glucose reading given after a blood-alcohol combination is tested. By decreasing the light reflected from the strip, the optical electrical interpretation of the Dextrostix is altered by alcohol such that there is an apparent increase in the glucose level as read by the eyetone meter.[Abstract] [Full Text] [Related] [New Search]