These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [The kinetic determination of glucose with the glucose dehydrogenase method using the Eppendorf automatic analyzer 5040 (author's transl)]. Author: Bruckner A. Journal: J Clin Chem Clin Biochem; 1980 May; 18(5):275-8. PubMed ID: 7430956. Abstract: A kinetic determination of glucose from blood, urine and spinal fluid is described with use of the new automatic analyzer ACP 5040 Eppendorf. The method uses glucose-dehydrogenase which converts glucose to gluconic acid. The NADH formed can be measured by the increase in absorbance at 334 nm. Our variation of test methodology gives good precision, accuracy and a high performance speed. There is a good correlation with the hexokinase-glucose-6-phosphate-dehydrogenase end point method.[Abstract] [Full Text] [Related] [New Search]