  • Title: An improved signal processor for the ultrasonic dimension gauge.
    Author: Sia JB, Pace DG, Osborne MW, Zanko MT, Kovzelove F.
    Journal: Cardiovasc Res; 1980 Aug; 14(8):490-4. PubMed ID: 7438151.
    Abstract:
    The flip-flop producing a pulse proportional to the transit time of an ultrasonic burst between two piezoelectric crystals was often reset by transmission artefacts in sonomicrometers. A design improvement was made. The insertion of two monostable vibrators and one nand gate corrected the resetting caused by transmission artefacts. The new design produced a flip-flop pulse that was proportional to the transit time of the ultrasonic burst through the aorta. Ascending aortic diameter was monitored in unanaesthetised dogs for 6 to 8 weeks using the new design.