These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Interlaboratory evaluation of four RIA kits for determination of plasma cortisol, with special reference to accuracy: influence of matrix in calibration standards. Author: Lantto O, Björkhem I, Blomstrand R, Kallner A. Journal: Clin Chem; 1980 Dec; 26(13):1899-902. PubMed ID: 7438443. Abstract: We used four RIA kits for determination of plasma cortisol to assay 12 plasma samples in four laboratories, and compared the results with those obtained with a highly accurate method based on isotope dilution/mass spectrometry. All kits (Farmos, NEN, Sorin, and Clinical Assays) showed 10-50% deviation from the comparison method in the normal range (200-700 nmol/L). The inter- and intralaboratory imprecision (CV) ranged between 8-27% and 5-11%, respectively. Possible explanations for this low accuracy are discussed. Retention of an intact biological matrix in the calibration standards seems to be one prerequisite for accuracy.[Abstract] [Full Text] [Related] [New Search]