Search MEDLINE/PubMed
Title: A miniature cadmium telluride detector module for continuous monitoring of left-ventricular function. Author: Hoffer PB, Berger HJ, Steidley J, Brendel AF, Gottschalk A, Zaret BL. Journal: Radiology; 1981 Feb; 138(2):477-81. PubMed ID: 7455133. Abstract: The authors describe a miniature cadmium telluride (CdTe) detector module for continuous monitoring of ventricular function using an equilibrium radionuclide blood-pool label. The detector and collimator are small, light, and suitable for direct attachment to the chest wall. Clinical studies in 18 patients using a prototype system demonstrated reasonably good correlation with left-ventricular ejection fractions (LVEF) determined by first-pass studies performed with a multicrystal scintillation camera (r = 0.74) and gated equilibrium studies performed with a computerized sodium iodide (Nal) probe (r = 0.76). The CdTe device may prove to be useful in patients in intensive and coronary care units as well as in ambulatory patients.[Abstract] [Full Text] [Related] [New Search]