These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Validation of use of the Nageotte hemocytometer to count low levels of white cells in white cell-reduced platelet components. Author: Moroff G, Eich J, Dabay M. Journal: Transfusion; 1994 Jan; 34(1):35-8. PubMed ID: 7505960. Abstract: BACKGROUND: Determination of the white cell (WBC) count in WBC-reduced platelet components requires methods that have a detection limit in the range of approximately 5.0 x 10(2) to 5.0 x 10(4) per mL. STUDY DESIGN AND METHODS: With a 50-microL Nageotte hemocytometer and bright-field microscopy (200x magnification), studies were conducted to develop and validate a method that could be used routinely with filtered and apheresis-harvested platelets. A 1-in-5 dilution of sample with a commercially available blood-diluting fluid was used because, with a lower (1-in-2) dilution, the observed number of WBCs was substantially less than the number expected at relatively high platelet counts (> 1.9 x 10(9)/mL). RESULTS: The observed and expected WBC counts in WBC-reduced platelet samples correlated well at levels between approximately 5 and 1100 WBCs per counting area (5.0 x 10(2)-1.1 x 10(5)/mL). At levels of more than 300 to 400 WBCs per counting area, accurate counts were obtained when 10 of the 40 rectangles were counted. CONCLUSION: These studies provide data to confirm that the 50-microL Nageotte hemocytometer can be used to accurately count low levels of WBCs in platelet components.[Abstract] [Full Text] [Related] [New Search]