These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Candidate reference method for determining serum creatinine by isocratic HPLC: validation with isotope dilution gas chromatography-mass spectrometry and application for accuracy assessment of routine test kits. Author: Thienpont LM, Van Landuyt KG, Stöckl D, De Leenheer AP. Journal: Clin Chem; 1995 Jul; 41(7):995-1003. PubMed ID: 7600702. Abstract: We present a candidate Reference Method for determining creatinine in serum, based on isocratic HPLC. The chromatographic column, with alkaline-treated aluminum oxide as stationary phase, is eluted with an equivolume mixture of methanol/acetonitrile containing 70 mL/L aqueous NaOH (10 mmol/L), and ultraviolet absorbance is detected at 240 nm. We investigated the ruggedness of the method and validated its performance with isotope dilution gas chromatography-mass spectrometry (ID GC-MS) for a set of 22 patients' sera and 10 commercially available lyophilized control materials. The mean deviation from ID GC-MS was +0.1% (range, -2.2% to +2.2%). The between-day CV, calculated from six independent measurements performed on three different days, was 0.9% (range 0.2% to 1.9%); the within-run CV was 0.8%. The total error of the method was < 3%. These performance characteristics make the method suitable for target-setting of quality-control materials and accuracy assessment of routine test kits. For the latter application, done with split-sample measurements of a panel of 83 patients' specimens, we used the Boehringer Mannheim enzymatic creatinine PAP test on the Hitachi 911, the Kodak Ektachem single-slide enzymatic creatinine test on the Ektachem 700, the Merck Jaffé kinetic assay on the Mega analyzer, and the Roche Jaffé kinetic test on the Cobas Mira.[Abstract] [Full Text] [Related] [New Search]