Title: SPECT imaging of fluorine-18. Author: Leichner PK, Morgan HT, Holdeman KP, Harrison KA, Valentino F, Lexa R, Kelly RF, Hawkins WG, Dalrymple GV. Journal: J Nucl Med; 1995 Aug; 36(8):1472-5. PubMed ID: 7629597. Abstract: UNLABELLED: The objective of this work was to determine the potential clinical usefulness of SPECT to image 511-keV annihilation photons. METHODS: A triple-headed gamma camera equipped with ultra-high-energy collimators was used to image 18F. Sensitivity measurements were carried out and the FWHM and FWTM were determined in air and for a unit-density scattering medium. Additionally, tomographic phantom studies were acquired to evaluate image quality. RESULTS: The sensitivities of the three cameras were, for all practical purposes, identical. At a source-to-collimator distance of 100 mm, the FWHM and FWTM were 13 and 29 mm, respectively. A tomographic phantom study demonstrated that spheres with a diameter of 20 mm were well resolved when filled with 18F activity and placed inside a water-filled phantom. CONCLUSION: The triple-headed SPECT camera in this investigation is a practical means of acquiring tomographic 18F images. The reconstructed slices were of sufficient quality to be of value in some clinical studies.[Abstract] [Full Text] [Related] [New Search]