  • Title: Loudness summation for two channels of stimulation in cochlear implants: effects of spatial and temporal separation.
    Author: McKay CM, McDermott HJ, Clark GM.
    Journal: Ann Otol Rhinol Laryngol Suppl; 1995 Sep; 166():230-3. PubMed ID: 7668649.
    Abstract:
    An experiment with four implantees with the Mini System 22 device was undertaken to measure the loudness summation across two channels of stimulation, with stimuli in which the current pulses were delivered alternately to each channel. The effects of varying spatial separation, temporal separation, and extent of stimulation were investigated. It was found that the absolute amount of summation varied among subjects, and was in general independent of electrode separation, except for a reduction at zero separation. Widening of the spatial extent of the stimulation did not have a consistent effect. There was a reduction in summation for all subjects at zero electrode separation when the time between the two pulses was increased from less than 1 millisecond to 2 milliseconds. In conclusion, loudness summation did not appear to be highly dependent on parameters that affect the spatial current spread in the cochlea. Further study of the effect of temporal parameters on loudness may help to quantify interaction between stimulation channels.
    [Abstract] [Full Text] [Related] [New Search]