These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Specific heparin preparations interfere with the simultaneous measurement of ionized magnesium and ionized calcium. Author: Lyon ME, Bremner D, Laha T, Malik S, Henderson PJ, Kenny MA. Journal: Clin Biochem; 1995 Feb; 28(1):79-84. PubMed ID: 7720231. Abstract: OBJECTIVE: To determine whether heparin anticoagulants used for analysis of whole blood ionized calcium would influence the measurement of ionized magnesium. METHODS: The effects of zinc heparin, lithium heparin, and electrolyte-balanced heparin on the simultaneous measurement of ionized magnesium and ionized calcium in serum were determined using ion selective electrodes. RESULTS: Time-dependent biases in ionized magnesium and calcium concentrations were apparent with zinc heparin but not with lithium or electrolyte-balanced heparin. Ionized magnesium and calcium concentrations were more significantly influenced by volume-dependent changes in zinc heparin potency than with lithium or electrolyte-balanced heparin. CONCLUSION: Zinc heparin produces a significant positive bias in the simultaneous determination of ionized magnesium and ionized calcium concentrations.[Abstract] [Full Text] [Related] [New Search]