These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Electrochemical determination of low blood lead concentrations with a disposable carbon microarray electrode.
    Author: Feldman BJ, D'Alessandro A, Osterloh JD, Hata BH.
    Journal: Clin Chem; 1995 Apr; 41(4):557-63. PubMed ID: 7720246.
    Abstract:
    Lead concentrations in blood samples typical of unexposed individuals are determined by using square-wave anodic stripping voltammetry at a disposable 287-element (15-35-microns diameter) carbon microarray electrode. Analysis of a series of low Pb-containing samples (blood [Pb] = 17 to 92 micrograms/L) by standard addition gives good accuracy (average bias vs graphite furnace atomic absorption spectroscopy = -4 micrograms/L) and precision (pooled SD = 7 micrograms/L). The signal/noise ratio is improved by about an order of magnitude over that found at a single carbon disk (10 microns diameter), resulting in a detection limit (for the microarray) of 5 micrograms/L for a 60-s deposition of sample.
    [Abstract] [Full Text] [Related] [New Search]