  • Title: Nitrate determination in biological fluids by an enzymatic one-step assay with nitrate reductase.
    Author: Bories PN, Bories C.
    Journal: Clin Chem; 1995 Jun; 41(6 Pt 1):904-7. PubMed ID: 7768010.
    Abstract:
    We report a simple, enzymatic, end point method for determining nitrate in serum and urine with use of nitrate reductase from Aspergillus sp. (EC 1.6.6.2). The decrease in absorbance at 340 nm as a result of the oxidation of beta-NADPH is measured. We used FAD as a supplementary electron carrier and added an internal standard to avoid interference from possible inhibitors of the enzymatic reaction. The method was linear from 5 to 200 mumol/L nitrate in serum. Within-run (n = 12) and total (n = 14 days) imprecisions (CV) were 5.1-7.7% and 6.2-9.8%, respectively, at 20-90 mumol/L nitrate in serum. Recoveries of added nitrate were 80-106%. The median (range) concentration in serum of 20 healthy subjects was 16 (0-42) mumol/L.