These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Reversed-phase high performance liquid chromatographic determination of cyproheptadine from urine by solid-phase extraction.
    Author: Kountourellis JE, Ebete KO.
    Journal: J Chromatogr B Biomed Appl; 1995 Feb 17; 664(2):468-71. PubMed ID: 7780605.
    Abstract:
    A simple, sensitive and reliable high-performance liquid chromatographic method for the determination of cyproheptadine in urine by solid-phase extraction (SPE) has been developed. The sample matrix was passed through a pre-conditioned C18 cartridge, washed with methanol-water solution (4:1) and eluted with methanol. The methanolic solution was evaporated to dryness, reconstituted with methanol and chromatographed using a C18 reversed-phase column. The mobile phase consisted of acetate buffer (constant ionic strength of 0.005 I)-methanol (56:44, v/v). Detection was performed at 254 nm with the sensitivity set at 0.002 AUFS. Concentrations as low as 50 ng/ml could be quantitatively determined by an external standard method and the overall recovery was found to be 76.16%, whereas the limit of detection was estimated as 15 ng/ml.
    [Abstract] [Full Text] [Related] [New Search]