These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Measurement of serum Lp(a) by COBAS MIRA using a latex immunoturbidimetric assay kit.
    Author: Fujita S, Sano T, Katayama Y.
    Journal: J Clin Lab Anal; 1994; 8(6):385-90. PubMed ID: 7869177.
    Abstract:
    We had an opportunity to test a reagent kit for serum lipoprotein (a) [Lp(a)] assay, which was based on the latex immunoturbidimetric assay method, and applied it to COBAS MIRA (Roche). Reproducibility of the assay procedure was 0.92-4.0% CV of three samples which contained 12.58-60.4 mg/dl of Lp(a). Minimum detectable concentration was 1.5-2.0 mg/dl. It was confirmed that interference of co-existing substances, i.e. hemoglobin, bilirubin, and intralipid, was negligible. No cross reactivity was seen with plasminogen. Correlation with a ELISA method was excellent. Frequency distribution of Lp(a) in healthy Japanese was similar to that found in white population.