These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Noninvasive monitoring of end-tidal CO2 via nasal cannulas in spontaneously breathing children during the perioperative period.
    Author: Tobias JD, Flanagan JF, Wheeler TJ, Garrett JS, Burney C.
    Journal: Crit Care Med; 1994 Nov; 22(11):1805-8. PubMed ID: 7956285.
    Abstract:
    OBJECTIVE: To determine the correlation between end-tidal CO2 and PaCO2 values measured via nasal cannulas in spontaneously breathing children during the perioperative period. DESIGN: Prospective evaluation. SETTING: Pediatric intensive/intermediate care unit in a tertiary care referral center. PATIENTS: Thirty postoperative surgical and trauma patients aged < or = 18 yrs (average age 7.8 yrs [range 6 months to 16 yrs] and average weight 28.3 kg (range 8.5 to 69). MEASUREMENTS AND MAIN RESULTS: Spontaneously breathing, nonintubated patients with an arterial cannula in place were selected for study. End-tidal CO2 was sampled from nasal cannulas by a sidestream aspirator and was estimated by infrared spectroscopy. The difference between PaCO2 and end-tidal CO2 was compared using linear regression analysis. A total of 55 blood gas measurements were obtained on the 30 patients. The PaCO2 to end-tidal CO2 gradient was < or = 4 torr in 54 of the 55 samples. The mean PaCO2 was 39.5 +/- 3.3 torr (5.27 +/- 0.44 kPa) with a mean end-tidal CO2 value of 39.7 +/- 3.8 torr (5.29 +/- 0.51 kPa). Linear regression analysis of arterial vs. end-tidal CO2 yielded a slope of 0.992 and p = .0001. CONCLUSIONS: End-tidal CO2 measurement by infrared spectroscopy provided an accurate estimation of PaCO2 in this patient population. Its use may limit the need for invasive monitoring and/or repeated arterial blood gas analysis.
    [Abstract] [Full Text] [Related] [New Search]