These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Usefulness of a third Holter lead for detection of myocardial ischemia.
    Author: Lanza GA, Mascellanti M, Placentino M, Lucente M, Crea F, Maseri A.
    Journal: Am J Cardiol; 1994 Dec 15; 74(12):1216-9. PubMed ID: 7977093.
    Abstract:
    Two-channel ambulatory electrocardiographic (ECG) monitoring is a useful method for detecting transient myocardial ischemia in patients with coronary artery disease. However, the monitoring of only 2 leads may fail to detect a significant number of ischemic episodes. In this study, the additional diagnostic value of a third bipolar chest lead was evaluated by recording a simultaneous 12-lead electrocardiogram and a 3-channel ambulatory electrocardiogram during exercise testing in 223 patients (aged 63 +/- 10 years) with proved or suspected coronary disease. Leads CM5, CM3, and an inferior lead (Y-modified or CMf) were monitored on the ambulatory electrocardiogram. Diagnostic ST-segment depression on the standard electrocardiogram was detected in 98 patients (44%), 94 (96%) of whom also had diagnostic ST-segment changes on the ambulatory electrocardiogram. Two additional patients had ST-segment depression only on the ambulatory electrocardiogram (both in lead CM5). Maximal ST-segment depression and duration of ischemia detected on standard and ambulatory ECG leads were similar in the 94 patients in whom ST-segment changes were detected on both types of ECG monitoring. CM5 was the single lead with the highest sensitivity (89%) in detecting myocardial ischemia. The addition of CM3 to CM5 increased sensitivity to 91%, and the addition of an inferior lead to CM5 increased sensitivity to 94%, particularly improving the detection of isolated inferior myocardial ischemia. The combination of all 3 ambulatory ECG leads had a sensitivity of 96%, an improvement of only 2% compared with the best combination of 2 leads (i.e., CM5 +/- inferior lead).(ABSTRACT TRUNCATED AT 250 WORDS)
    [Abstract] [Full Text] [Related] [New Search]