These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: High flow closed circuit anaesthesia. Author: Verkaaik AP, Van Dijk G. Journal: Anaesth Intensive Care; 1994 Aug; 22(4):426-34. PubMed ID: 7978208. Abstract: We present an automatic closed circuit anaesthesia ventilator designed for routine clinical use. The ventilator combines the benefits of high flow systems and true closed circuits, without their disadvantages. The system can be used with any FiO2, with air or nitrous oxide as carrier gas. Servo controlled delivery of modern volatile anaesthetics is regulated on end-tidal value. The time constant for increase or decrease of concentrations is only a few minutes. There is no need to open the system at any time, nor is it necessary to increase the fresh gas flow. An automatic flush procedure prevents accumulation of unwanted gases. Operation is as easy as contemporary non-closed circuit ventilators. With this machine, closed circuit anaesthesia is possible from the beginning to the end of the procedure.[Abstract] [Full Text] [Related] [New Search]