These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Comparison of Swan neck catheter with Tenckhoff catheter for CAPD. Author: Hwang TL, Huang CC. Journal: Adv Perit Dial; 1994; 10():203-5. PubMed ID: 7999828. Abstract: To decrease the incidence of exit-site infection, the Swan neck catheter has been used since 1991. Twenty-six patients received Swan neck catheter implants over the past three years. Fifteen patients suffered from exit-site infections and 2 of them developed tunnel infections. The incidence of exit-site or tunnel infections in the Swan neck group revealed no difference from those patients receiving Tenckhoff tube implants (n = 154). Two patients in the Swan neck group and 8 in the Tenckhoff group had tube migrations. There was no difference between the two groups. No patient in the Swan neck group suffered from cuff extrusion or pericatheter leakage, while 9 in the Tenckhoff group suffered from cuff extrusions and 5 had pericatheter leakages. The cumulative catheter survival rates at one, two, and three years in the Swan neck group were 90.9%, 80.2%, and 67.9%, while those in the Tenckhoff group were 84.7%, 67.9%, and 54%. We conclude that the patients receiving Swan neck catheter implantations had a lower incidence of cuff extrusion and pericatheter leakage, and better catheter survival, though the incidence of exit-site/tunnel infection and tube migration was no better than in the Tenckhoff catheter group.[Abstract] [Full Text] [Related] [New Search]