These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: [18 months experience with a new single needle cytapheresis system (Fresenius AS 104 SN)]. Author: Zeiler T, Zingsem J, Weisbach V, Zimmermann R, Wittmann G, Eckstein R. Journal: Infusionsther Transfusionsmed; 1994 Apr; 21(2):91-5. PubMed ID: 8019115. Abstract: BACKGROUND: Since October 1991 the third-generation cell separator Fresenius AS 104 offers a single-needle (SN) option for thrombocytapheresis. Here we present our first 18-month experience with this new single-needle system. MATERIALS AND METHODS: We performed 395 thrombocytaphereses in 225 donors. The influence of blood flow, cycle volume, and interface position on efficacy and leukocyte contamination was evaluated. RESULTS: 395 thrombocytaphereses were performed with an average thrombocyte yield of 3.00 +/- 0.69 x 10(11) platelets and a leukocyte contamination of 1.66 +/- 3.1 x 10(8) per concentrate. Blood flow rates of 50 ml/min (vs. 60 ml/min) and an interface position of 6:2 (vs. 7:1) resulted in a lower leukocyte contamination, but without statistical significance. CONCLUSION: The SN program of the AS 104 provides high thrombocyte yields and less leukocyte contamination than other SN systems, but 77% of the concentrates still contain more than 0.5 x 10(8) leukocytes.[Abstract] [Full Text] [Related] [New Search]