These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Pin removal after in-situ pinning for slipped capital femoral epiphysis. Author: Bellemans J, Fabry G, Molenaers G, Lammens J, Moens P. Journal: Acta Orthop Belg; 1994; 60(2):170-2. PubMed ID: 8053316. Abstract: We retrospectively reviewed 28 pin-removal procedures in 20 patients with slipped capital femoral epiphysis, who were all treated by in-situ pinning with Kowles pins. Failure to extract the pins occurred in 8 cases (28.6%), and in another 8 cases (28.68%), pin retrieval was only possible after major excavation of the lateral femoral cortex. In view of these data we do not recommend routine removal of Knowles pins used for in-situ pinning of slipped capital femoral epiphysis.[Abstract] [Full Text] [Related] [New Search]