These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of ascorbic acid in human plasma by high-performance liquid chromatography with electrochemical detection using a hydroxyapatite cartridge for precolumn deproteinization.
    Author: Iwase H, Ono I.
    Journal: J Chromatogr B Biomed Appl; 1994 May 13; 655(2):195-200. PubMed ID: 8081465.
    Abstract:
    Ascorbic acid (AA) was determined in human plasma by high-performance liquid chromatography with electrochemical detection using a hydroxyapatite cartridge for plasma deproteinization. The proposed method is simple, rapid (deproteinization time, within 1 min; analysis time, ca. 15 min), sensitive [detection limit, ca. 240 ng/ml plasma (at a signal-to-noise ratio of 2:1)], highly selective and reproducible [relative standard deviation, ca. 2.8% (n = 3)]. The calibration graph for AA was linear in the range 0.1-10 ng per injection (20 microliters). The recovery of AA was over 90% by the standard addition method.
    [Abstract] [Full Text] [Related] [New Search]