  • Title: Cardiac troponin-T immunoassay for diagnosis of acute myocardial infarction.
    Author: Wu AH, Valdes R, Apple FS, Gornet T, Stone MA, Mayfield-Stokes S, Ingersoll-Stroubos AM, Wiler B.
    Journal: Clin Chem; 1994 Jun; 40(6):900-7. PubMed ID: 8087984.
    Abstract:
    We evaluated the analytical and clinical performance of an immunoassay for cardiac troponin T (cTnT). Within-run and total imprecision ranged from 1.6 to 11.3%. The sensitivity and linear range was 0.015 and 13 micrograms/L, respectively. Frozen samples were stable for at least 8 weeks. No interferences were seen with lipids or bilirubin (total and conjugated). Hemoglobin caused a negative bias at concentrations > 4 g/L. Heparinized plasma showed a 6% negative bias compared with serum. The clinical utility of cTnT was compared with that of creatine kinase (CK)-MB (mass assay). The sensitivity of cTnT measurements from 63 patients with acute myocardial infarction (AMI) (cTnT cutoff 0.1 microgram/L) was 60% at 0-3 h, 59% at 3-6 h, 94% at 6-9 h, 90% at 9-12 h, 99% at 12-24 h, 92% at 24-48 h, 83% at 48-72 h, and 100% at 72-96 h. Corresponding results for CK-MB (cutoff 5.0 micrograms/L and 2.5% relative index) were 45%, 64%, 82%, 97%, 87%, 81%, 54%, and 59%. The specificity of the markers from 49 non-AMI patients was 46% and 79% for cTnT and CK-MB, respectively. We show that CK-MB is more specific for diagnosis of AMI, and propose that cTnT is more sensitive to myocardial injury.
    [Abstract] [Full Text] [Related] [New Search]