These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Cardiac troponin-T immunoassay for diagnosis of acute myocardial infarction. Author: Wu AH, Valdes R, Apple FS, Gornet T, Stone MA, Mayfield-Stokes S, Ingersoll-Stroubos AM, Wiler B. Journal: Clin Chem; 1994 Jun; 40(6):900-7. PubMed ID: 8087984. Abstract: We evaluated the analytical and clinical performance of an immunoassay for cardiac troponin T (cTnT). Within-run and total imprecision ranged from 1.6 to 11.3%. The sensitivity and linear range was 0.015 and 13 micrograms/L, respectively. Frozen samples were stable for at least 8 weeks. No interferences were seen with lipids or bilirubin (total and conjugated). Hemoglobin caused a negative bias at concentrations > 4 g/L. Heparinized plasma showed a 6% negative bias compared with serum. The clinical utility of cTnT was compared with that of creatine kinase (CK)-MB (mass assay). The sensitivity of cTnT measurements from 63 patients with acute myocardial infarction (AMI) (cTnT cutoff 0.1 microgram/L) was 60% at 0-3 h, 59% at 3-6 h, 94% at 6-9 h, 90% at 9-12 h, 99% at 12-24 h, 92% at 24-48 h, 83% at 48-72 h, and 100% at 72-96 h. Corresponding results for CK-MB (cutoff 5.0 micrograms/L and 2.5% relative index) were 45%, 64%, 82%, 97%, 87%, 81%, 54%, and 59%. The specificity of the markers from 49 non-AMI patients was 46% and 79% for cTnT and CK-MB, respectively. We show that CK-MB is more specific for diagnosis of AMI, and propose that cTnT is more sensitive to myocardial injury.[Abstract] [Full Text] [Related] [New Search]