These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Rapid and sensitive assay of dobutamine in plasma by high-performance liquid chromatography and electrochemical detection. Author: Husseini H, Mitrovic V, Schlepper M. Journal: J Chromatogr; 1993 Oct 22; 620(1):164-8. PubMed ID: 8106586. Abstract: A sensitive and specific high-performance liquid chromatographic method with electrochemical detection was developed for measuring dobutamine in human plasma samples. Following an external standard method, 0.1 ml of EDTA-glutathione plasma was diluted on ice with 0.2 ml of a 5% trichloracetic acid solution. The mixture was centrifuged, filtered, and 30 microliters were injected. Assessment was done by electrochemical detection. The assay was linear from 1 to 400 ng/ml plasma. For determination of dobutamine we also used a liquid-liquid extraction method routinely applied for plasma catecholamines. Liquid-liquid extraction requires application of 100-1000 microliters of plasma. The standard curve was linear from 0.1 to 600 ng/ml. Absolute recovery of dobutamine was 90 +/- 3% with the liquid-liquid extraction procedure and 91 +/- 3% with the protein precipitation method. For both methods dobutamine was separated on Nova-Pak C18 columns. The mobile phase used was 0.1 molar phosphate buffer-acetonitrile (80:20, v/v) with 1-octanesulfonic acid and triethylamine as ion-pair reagents. The pH was adjusted to 2.7.[Abstract] [Full Text] [Related] [New Search]