These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: In vivo evaluation of an experimental dental HA-Ti implant.
    Author: Gatti AM, Chiarini L, Monari E, Tanza D.
    Journal: Med Prog Technol; 1993; 19(3):139-44. PubMed ID: 8127279.
    Abstract:
    A new design of Titanium dental implant, coated with Titanium oxide and with a Hydroxyapatite apex, was experimented in a Beagle dog's jaw for one year to evaluate the biocompatibility of the materials used and the biomechanical efficiency of the design. The explanted prosthesis and the surrounding bone were embedded in methylmethacrylate resin and the sections observed under Scanning Electron Microscopy (SEM), with an X-ray microprobe to check the interface with bone and the occurrence of ionic release. The results showed that the pin was well accepted by the bone. New bone grew on the metal surface developing a mechanical bond with the prosthesis. No ionic release of the metal was seen.
    [Abstract] [Full Text] [Related] [New Search]