These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: [Findings of scanning microscopy studies and gas-transfer results in long-term perfusions using membrane oxygenators in animal experiment].
    Author: Herzer JA, Falke K, Schulte HD, Krian A, Rosenbauer KA, Bircks W.
    Journal: Langenbecks Arch Chir; 1975; Suppl():87-92. PubMed ID: 813074.
    Abstract:
    Twelve hours partial extracorporeal circulation using the GEDL was carried out in sheep with experimental pulmonary insufficiency. The maximal O2-transfer was 53,8 ccm/min/m2 membrane surface at blood flows of 2 and 3 1/min. The CO2-transfer was nearly 30 ccm/min/m2 if the pCO2 was within the normal range. There was evidence, that the O2-transfer decreases during prolonged perfusion. Scanning-electron-microscopic studies revealed a thickening of the membrane with protein and cell-fragments which were thought to be the reason for the observed decrease in O2-transfer.
    [Abstract] [Full Text] [Related] [New Search]